Teledyne 2000RS Operating Instructions Manual page 78

Thermal conductivity analyzer
Table of Contents

Advertisement

TCD Analyzer
 "Connected" if the 2000RS board is connected to the external
Profibus master PLC.
"No Connection" if the 2000RS board Profibus
communication has failed to communicate with the external
with Profibus Master PLC.
4.5.10.1
P
ROFIBUS
The Profibus task will receive remote commands from a master PLC.
These are the System module input registers:
Address
Byte Location
0x00
The System output module registers are organized like this:
Group
Address
0x00
Gas
0x01
Concentration
Value
0x02
(0x00 - 0x03)
0x03
0x04
0x05
Reserved
(0x04 - 0x07)
0x06
0x07
Teledyne Analytical Instruments
T
ASK
Variable
Analyzer
0x00
Command
Bit
Variable
Number
concentration
(LSB)
concentration
(MSB)
Reserved
Operation
Variable
description
0x01 - Analyze
0x08 - Span
0x10 - Zero
0x20 - ESC
Variable description
Gas
Current gas
concentration values
updated.
Gas
Reserved
65