Auto Mode Spanning - Teledyne 2000RS Operating Instructions Manual

Thermal conductivity analyzer
Table of Contents

Advertisement

TCD Analyzer
4.8.2.1 A
M
UTO
ODE
Press Span to enter the span function.
Menu screen:
The screen that appears allows you to select whether the span
calibration is to be performed automatically or manually.
This screen also allows the operator to set the time the analyzer
should be held in the span mode after the readings of the analyzer settle.
Five minutes is the default, but it could be adjusted anywhere from 1 to
60 minutes by user.
This screen is also used to enter the span concentration of the
calibration gas.
Span Cal configuration screen:
Teledyne Analytical Instruments
S
PANNING
Operation