TCD Analyzer
Span concentration:
To enter or check the Span Concentration of the calibration gas
touch the corresponding edit box to bring up the numerical pad.
Enter the span concentration using the same procedure as for the
cal holding time described above. When you have finished typing in the
concentration of the span gas you are using, press ENT to save the value
or the back arrow to discard the entry and dismiss the numerical pad.
Start the Calibration:
To begin the Span calibration, touch the corresponding box on the
Span Cal screen.
Teledyne Analytical Instruments
Operation
95