TCD Analyzer
Calibration status bar will display a message of "Calib is starting..."
and then system will wait till the hold time completes. During the time
"Calib progress. (48)" message will appear in the status bar.
Zero calibration auto mode run screen contains the following
options (Refer above image):
1. Calibration is in progress icon
calibration mode. Calibration is finished or an error occurs
when progress icon stops refreshing,
2. Gas % or ppm values
3. Slope values
4. Calibration status bar
5. Temperature status
The calibration status bar, below "Slope" is counting down the
timer set in "Calib holding time". Once the slope value is settled and
user set "Calib holding time" duration is completed (timer counter is
down to zero) then system will start the coarse setting process. The
icons on the screen switches to the Coarse setting value as software tries
to adjust reading down to zero.
Teledyne Analytical Instruments
Operation
. It only appears in AUTO
89