Maintenance
Replacing the battery periodically ensures battery operation
when needed. A periodic battery check is also recommended.
The interval for this check is at the user's discretion.
Depending on how the monitor is used, the user may wish to
perform a battery check as often as every 6 months. This check
should also include inspecting the battery's physical condition.
Any batteries showing leakage or corrosion should be replaced.
Qualified service personnel should check the battery following
instructions given in the N-200 service manual. Regardless of
periodic test results, Nellcor recommends battery replacement
every 2 years. Replacement batteries can be purchased through
Nellcor's Technical Services Department or your local Nellcor
representative.

DETERMINING SOFTWARE VERSION

To determine the N-200 software version, press the HIGH SAT
and AUDIO ALARM OFF buttons. A two-digit version number
appears in the PULSE RATE display. For example, version 2.7
appears as 27. It may be necessary to turn the control knob
until 0 appears in OXYGEN SATURATION display. To
determine the powerbase software version, press the HIGH
SAT and AUDIO ALARM OFF buttons again, and turn the
control knob to the left until Pb appears in the OXYGEN
SATURATION display; the version number appears in the
PULSE RATE display.

TECHNICAL ASSISTANCE

For technical information or to order parts or a service manual,
contact Nellcor's Technical Services Department or your local
Nellcor representative. The service manual includes circuit
diagrams and a spare parts list, which may be required by
qualified personnel when servicing the N-200.