The Print-mode indicator (page icon) flashes at the
medium rate throughout the print mode.
When the print function is complete, the display returns
to the normal mode. When NPB-40 times-out before the
Data button is pressed, the display returns to the normal
mode.
The Menu button is disabled during the print mode.
If there is no communication with the printer for
30 seconds, the print mode is aborted and the display
returns to the normal mode.
Summary report data is purged after it is printed or on
time-out due to printer communication failure. Printing
or time-out due to printer communication failure does not
purge snap-shot or sensor-event data; this data is purged
from the NPB-40 memory at power off. Sensor-event
data is retained in the sensor and loads/reloads to the
NPB-40 memory at power on.
Alarm annunciation and patient monitoring is disabled
during active printing (when the NPB-40 is
communicating with printer).

Printing Data

Discussion

The NPB-40 prints:
summary report, if available
stored snap-shot data, if available
sensor-event data, if available