OXIMAX N-65 Operation
Print Mode
If you press the Data button in the print menu when no
data is stored, and the device is not connected to a
patient, an invalid key tone sounds and the monitor
transitions to normal mode.
The Print-mode indicator (page icon) flashes at the
medium rate throughout print mode.
When the print function is complete, the display returns
to the normal mode. If the monitor times-out before the
Data button is pressed, the display returns to the normal
mode.
The Menu button is disabled during print mode.
If there is no communication with the printer for
30 seconds, print mode is aborted and the display returns
to normal mode.
After Printing Data
The summary report data is purged after it is printed or
on time-out due to printer communication failure.
Snap-shot or sensor-event data is not purged after
printing or time-out due to printer communication failure.
Instead, it is purged from the monitor's memory at power
off. Sensor-event data is retained in the sensor and loads/
reloads to the monitor memory at power on.
Alarm annunciation and patient monitoring is disabled
during active printing when the monitor is
communicating with printer.