Perform C-Echo Test
Use this function when you want to check to see if the system is communicating with a particular
network host.
For mobile units, the unit must be connected to the network for this test to be of any value.
1. Select a host from the Network Hosts list.
2. Press [C-ECHO TEST].
A message appears to notify you if the test passed or failed.
A "passed" message means that the network host is working and that you can retrieve exams
from it or transfer exams to it.
Failed C-Echo Test
A "failed" message means that the system could not contact the network host.
Perform the following tasks to resolve the problem:
•
Check that the unit is connected to the network.
•
Try connecting the unit to a different network port. There may be a problem with the port itself.
•
For units with wireless connection:
– Check the signal strength
Remove any metal that is covering the antenna (e.g, lead apron, clip board)
Move the unit to another area with better reception
– Try software reset to establish the wireless connection
•
Re-try the test at a later time. The host may be temporarily unavailable.
•
Confirm the host configuration on the Edit screen.
•
If the problem persists, contact your technical support group or system administrator.
Printers
Follow this process to access the Network Connections screen.
1. On the Worklist screen, press [UTILITIES].
The System-System screen appears.
2. Press [NETWORK CONNECTIONS].
The Network Connections screen (Figure A7-5) appears.
Definium AMX 700 X-Ray System
Operator Manual
5161515-1EN
Rev. 6 (10 February 2008)
© 2008 General Electric Company. All rights reserved.
Chapter A7: Set Preferences
A7-7