Perform C-Echo Test; Printers - GE Definium AMX 700 Operator's Manual

X-ray system
Table of Contents

Advertisement

Perform C-Echo Test

Use this function when you want to check to see if the system is communicating with a particular
network host.
For mobile units, the unit must be connected to the network for this test to be of any value.
1. Select a host from the Network Hosts list.
2. Press [C-ECHO TEST].
A message appears to notify you if the test passed or failed.
A "passed" message means that the network host is working and that you can retrieve exams
from it or transfer exams to it.
Failed C-Echo Test
A "failed" message means that the system could not contact the network host.
Perform the following tasks to resolve the problem:
Check that the unit is connected to the network.
Try connecting the unit to a different network port. There may be a problem with the port itself.
For units with wireless connection:
– Check the signal strength
Remove any metal that is covering the antenna (e.g, lead apron, clip board)
Move the unit to another area with better reception
– Try software reset to establish the wireless connection
Re-try the test at a later time. The host may be temporarily unavailable.
Confirm the host configuration on the Edit screen.
If the problem persists, contact your technical support group or system administrator.

Printers

Follow this process to access the Network Connections screen.
1. On the Worklist screen, press [UTILITIES].
The System-System screen appears.
2. Press [NETWORK CONNECTIONS].
The Network Connections screen (Figure A7-5) appears.