5. Repeat process until all exams are selected for all procedures.
6. Press [ACCEPT] when finished.
The Select Protocols screen closes.
The Acquisition screen appears.
[CLOSE] removes the selections and returns you to the Worklist.
7. Refer to

Conduct a Digital Exam

exam.
Conduct a Digital Exam
This section describes the adjustments required when conducting a digital exam.
Follow this process to conduct an exam using the GE Portable DR Detector.
NOTE: If you need to interrupt the exam and resume it at a later time, press the [SUSPEND] button. You
will be returned to the Worklist.
1. If multiple procedures were selected from the Worklist, select the procedure from the exam drop-
down list.
2. Select the Exam and View to perform from the protocol list.
3. Select the Patient Size. The system default is Medium Adult.
NOTE: To optimize processing for the best image quality, Patient Size should be confirmed for each view.
Available options are:
– Medium Pediatric (toddlers to 5 years old)
– Small Adult
– Medium Adult
– Large Adult
WARNING:
It is critical to select the proper patient size on the Acquisition screen. The incorrect Patient
Size may result in an unnecessarily large radiation dose or multiple exposures.
4. Select the GE Portable DR Detector receptor, if necessary.
Figure A4-6 Receptors: GE Portable DR Detector selected
Definium AMX 700 X-Ray System
Operator Manual
(p. A4-9) or
Conduct a Cassette Exam
5161515-1EN
Rev. 6 (10 February 2008)
© 2008 General Electric Company. All rights reserved.
Chapter A4: Image Acquisition
(p. A4-10) to continue the
A4-9