Performing A C-Echo Test - GE Healthcare Optima XR200amx Installation Manual

Hide thumbs Also See for Healthcare Optima XR200amx:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5336113-1EN, R
IRECTION
Allow this network host to send images
to the (system name)
This network host accepts multiple
frame image.
Perform de-identification (anonymous
patient images) when sending to this
network host.
Apply "Burn-On-Send" to images when
sending to this network host.
Send MPPS N-Create and N-Set
notification to this network host.
Make this network host the HIS/RIS
source. (Only one host can be
designated as the HIS/RIS source.)
DR Images sent as CR Modality
9.
10. Repeat this process for each network host device the site wants to add.

4.4.1.2 Performing a C-Echo test

Use this function to determine if the system is communicating with a particular network host.
10
EVISION
Table 4-13 Add Network Host – Preferences tab description
Function
Press SAVE.
Chapter 4 Configuration
Description
Available only on the Optima XR220amx or Optima
XR200amx with digital upgrade system.
Allows the host to send images to the system.
Available only on the Optima XR220amx or Optima
XR200amx with digital upgrade system.
Allows multiple frame images to be sent to this
network host.
Available only on the Optima XR220amx or Optima
XR200amx with digital upgrade system.
Automatically de-identifies (anonymizes) any
images that the system sends to this host.
Available only on the Optima XR220amx or Optima
XR200amx with digital upgrade system.
Burns the VOI LUT (Look-Up Tables) into the
DICOM header to be displayed by PACS. Leaving
this option unselected sends all available VOI LUTs
to the DICOM header for PACS to query and apply.
Note: PACS should be configured to read the first
VOI LUT for proper display of images on the
Acquisition workstation.
Available only on the Optima XR220amx or Optima
XR200amx with digital upgrade system.
This node acts as the Destination for receiving the
MPPS NCreate & N-Set Notification. When
configured for MPPS, the System sends information
like which exam is in progress, when a Study is
completed, how many images were acquired, and
what was the radiation dose to which the patient
was exposed during that session, etc.
Designates the host as the DICOM Worklist
provider. Defining the Radiology Information
System (RIS) and Hospital Information System
(HIS) host allows patient worklists to be downloaded
from those networks to the system.
Note: Only one HIS/RIS source may be designated
on the system. Selecting this option will de-select
any other hosts as the HIS/RIS source.
Available only on the Optima XR220amx or Optima
XR200amx with digital upgrade system.
Options are Yes or No.
O
XR200
/XR220
PTIMA
AMX
I
M
NSTALLATION
ANUAL
Page 69
AMX

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Healthcare optima xr220amx

Table of Contents