Qualified Service Available

GE Healthcare and its associates maintain a worldwide organization of service personnel specially
trained on medical x-ray equipment. Consult your GE representative to find out more about all the
available service programs.

Maintenance Schedule

Table 7-1 Periodic Maintenance Schedule
Maintenance/Checks
Replace counterweight
cable
Inspect counterweight cable
Verify no movement
between collimator and tube
unit
Extension arm inspection
Collimator general
inspection and adjustment
Filament and rotor interlocks
X-ray tube anode rotation
Column and carriage
assembly inspection
mAs accuracy
kVp accuracy
Operator displays
Cross hair
Collimator alignment
Field size indicator
Rotational detent
Verify all cable ties are tight
and secure
Definium AMX 700 X-Ray System
Operator Manual
Every 5 years
Every year (Refer to Service Documentation CD-ROM 5161519-200.)
Every 12 months, verify that there is no movement between the
collimator and X-ray tube. If any movement is detected, you MUST
remove the collimator and inspect all related fasteners. Refer to
Service Documentation CD-ROM 5161519-200 for additional
information.
Every year (Refer to Service Documentation CD-ROM 5161519-200.)
Every year (Refer to Service Documentation CD-ROM 5161519-200.)
Every year (Refer to Service Documentation CD-ROM 5161519-200.)
Every year (Refer to Service Documentation CD-ROM 5161519-200.)
Every year (Refer to Service Documentation CD-ROM 5161519-200.)
Every year (Refer to Service Documentation CD-ROM 5161519-200.)
Every year (Refer to Service Documentation CD-ROM 5161519-200.)
Every year (Refer to Service Documentation CD-ROM 5161519-200.)
Every year (Refer to Service Documentation CD-ROM 5161519-200.)
Every year (Refer to Service Documentation CD-ROM 5161519-200.)
Every year (Refer to Service Documentation CD-ROM 5161519-200.)
Every year (Refer to Service Documentation CD-ROM 5161519-200.)
Every year (Refer to Service Documentation CD-ROM 5161519-200.)
5161515-1EN
Rev. 6 (10 February 2008)
© 2008 General Electric Company. All rights reserved.
Chapter 7: Maintenance and Service
Frequency
7-2