Good Operating Practices - GE Definium AMX 700 Operator's Manual

X-ray system
Table of Contents

Advertisement

CAUTION:
THIS SYSTEM IS NOT DESIGNED FOR USE IN CLOSE PROXIMITY TO A DEFIBRILLATOR!
Never use a defibrillator on a patient that remains in contact with the digital detector or
any part of the mobile x-ray system. This system must be treated as a conductive surface
and moved well away from a patient before defibrillation is attempted. If any part of the
mobile x-ray system remains in contact with a patient when the defibrillator is
discharged, voltage may be conducted through the patient's body and into the system.
This may be hazardous to anyone who may come in contact with the system, and could
damage the detector.
Always consult the instructions for use of any defibrillator that may be used on a patient
being imaged by this digital mobile x-ray system.
CAUTION:
If the GE Portable DR Detector is used in a fixed installation, it is the User's responsibility
to provide the means for audio and visual communication between the Operator and the
patient.
CAUTION:
If you suspect any electromagnetic interference affecting or caused by the unit, call
service. Portable and mobile RF communications equipment can affect medical electrical
equipment.
CAUTION:
This equipment is not rated for use in the presence of flammable gases.

Good Operating Practices

Wear a lead apron while performing an x-ray exam.
Step back at least 6 feet (1.8 meters) from the tube or to the full extension of the hand switch cord
before making an exposure.
Always use the proper field sizes and technic factors for each procedure to minimize x-ray exposure
and produce the best diagnostic results.
Check the digital display carefully before making an exposure: verify that the selected technique is
the intended technique. Pay particular attention to the placement of the decimal point in the mAs
setting to insure that whole numbers are not mistaken for an intended mAs fractional number.
When x-raying bed patients, move them as far as possible from nearby patients.
Ask visitors to step outside the room during an exposure.
Use gonadal shields for patients whenever possible.
Be sure to read and follow the maintenance schedule outlined in the Maintenance and Service
section of this manual.
Definium AMX 700 X-Ray System
Operator Manual
5161515-1EN
Rev. 6 (10 February 2008)
© 2008 General Electric Company. All rights reserved.
Chapter 2: Safety and Regulatory