When To Perform Quality Tests; Detector Check; Qap; Quality Test Preparation - GE Definium AMX 700 Operator's Manual

X-ray system
Table of Contents

Advertisement

When to Perform Quality Tests

This section describes the recommended quality testing schedule and any other events that warrant
quality tests to be performed

Detector Check

Detector Check is used daily to check detector quality and anytime after an accidental drop.
After a detector drop, the system reports an error and displays a QAP alert icon (Figure A8-1) to indicate
that you should conduct a Detector Check test on the detector.
The Detector Check alert icon will remain until a Detector Check or QAP test is run.

QAP

QAP should be performed:
On a scheduled, weekly basis.
When the alert icon appears on the QAP button (Figure A8-1).
When there is a perceived loss of image quality.
NOTE: When scheduling QAP tests, plan 15-20 minutes to perform the test. One portion of the test requires
that the system be powered on and in an idle state (i.e., without exposures) for at least 10 minutes.
The 10 minute idle period is based upon the time from the last exposure prior to performing QAP
or system start up. The rest of the test requires approximately 5 minutes to complete.
Figure A8-1 QAP button with alert icon

Quality Test Preparation

Before starting quality tests, be sure to do the following:
Close or suspend any open exams
Close any exam being reviewed
Clear all objects from detector and beam path (QAP only)
Disconnect the unit from the charging outlet (exposures are not permitted while the unit is charging)
The QAP process begins when the QAP button is clicked. The button is located at the bottom of the
Worklist or Acquisition screen.
NOTE: QAP cannot be performed if an exam is open or being reviewed.
Definium AMX 700 X-Ray System
Operator Manual
Chapter A8: Quality Assurance Process and Maintenance
5161515-1EN
Rev. 6 (10 February 2008)
© 2008 General Electric Company. All rights reserved.
A8-2

Advertisement

Table of Contents
loading

This manual is also suitable for:

5151481-35151481-4

Table of Contents