Figure A8-6 Detector Check results screen

Failed Detector Check

If the system fails the Detector Check (Figure A8-7), call to schedule service. The system is operational,
although inspection and potential calibration are needed.
WARNING:
Image quality may be degraded if you continue to use a detector that has failed the
Detector Check tests.
Figure A8-7 Failed Detector Check results

Perform QAP

QAP involves tests requiring exposures with the flat field phantom. The GE Portable DR Detector QAP kit
includes a flat-field phantom (Figure A8-8) sized to fit the collimator.