Figure 3-3 Hand switch
Table 3-3 Hand switch button positions
Position
OFF
PREP
EXPOSE

TechSwitch

The TechSwitch option may be purchased for use with the Definium AMX 700. The remote TechSwitch
can be used instead of the standard wired hand switch.
Definium AMX 700 X-Ray System
Operator Manual
The OFF position is when no pressure is applied to the Prep/Expose button
on top of the hand switch.
PREP is the next position on the hand switch. When it is partially pressed, it
brings the rotor up to speed and heats the filament. PREP also checks the
system interlocks and verifies the system is ready to make an exposure.
The button returns to the OFF position if released.
The EXPOSE position is when the button on the hand switch is fully
depressed. This produces X-rays that are recorded. Release the Prep/
Expose button after the exposure is completed.
5161515-1EN
© 2008 General Electric Company. All rights reserved.
Chapter 3: Hardware Overview
Description
Rev. 6 (10 February 2008)
1. PREP
2. EXPOSE
3. Collimator Lamp Switch
3-7