Export Dei Log; Exam - GE Definium AMX 700 Operator's Manual

X-ray system
Table of Contents

Advertisement

3. Repeat steps 1 and 2 for all applicable anatomical categories.
4. Press [SAVE] when finished to apply the changes and close the screen.
[CLOSE] closes the screen without saving the changes.
Enable or Disable Technical Mode
Technical Mode is a special setting that configures the system for image quality testing and detector
calibration. When Technical Mode is enabled, DEI is calculated using the central area of the image (512
pixels x 512 pixels) regardless of the imaged anatomy.
IMPORTANT!:
Do not acquire exam images when Technical Mode is enabled. The use of Technical Mode
is reserved for image testing and calibration purposes only.
The checkbox (Figure A7-21) indicates if Technical Mode is enabled. Check or uncheck the box as
appropriate and press [SAVE] to apply the change and close the screen.
Figure A7-21 Technical Mode checkbox

Export DEI Log

The DEI log records the exposure levels of images. Medical Physicists in your facility or GE Service
Personnel use the log to determine if there is a problem with the detector and to determine what
adjustments to make to correct exposure errors.
Follow this process to export the DEI log onto a CD.
1. If necessary, connect the external CD drive to the unit.
2. Insert a blank CD-R disc.
3. From the Worklist screen, press [UTILITIES].
4. Select Preferences > DEI.
5. Press the DEI Export Log [EXPORT] button.
6. Remove the disc when export is complete.

Exam

There are currently no preferences to configure on this screen.
Definium AMX 700 X-Ray System
Operator Manual
5161515-1EN
Rev. 6 (10 February 2008)
© 2008 General Electric Company. All rights reserved.
Chapter A7: Set Preferences
A7-30