13. Review the Status (far right column) of each test.
If all tests PASS: QAP is complete. Click [EXIT] to return to the Worklist screen and continue work.
If some tests FAIL: Refer to
Figure A8-15 QAP Results screen

Failed QAP

In the event of failed QAP tests, wait 10 minutes and repeat the QAP procedure to confirm the failure.
Check the items listed below when repeating the tests:
Check collimator blades. Make sure that they are fully open and not in the field of view.
Check that the phantom is correctly positioned.
Check that no other objects are in the field of view.
If a single failure is confirmed, call to schedule service. The system is operational, although inspection
and potential calibration are needed.
If multiple failures are confirmed, image quality may be affected; cease use of the unit and call for
immediate service.
Definium AMX 700 X-Ray System
Operator Manual
Chapter A8: Quality Assurance Process and Maintenance
Failed QAP
(p. A8-10) for more information.
5161515-1EN
Rev. 6 (10 February 2008)
© 2008 General Electric Company. All rights reserved.
A8-10