Detector Connector Pins - Teledyne Foxy R1 Instruction Manual

Hide thumbs Also See for Foxy R1:
Table of Contents

Advertisement

Connector
a. Input is +3.3 to 5 VDC or contact closure, active low, 30 ms minimum, 5 VDC maximum.
b. Output is open collector, active low, 15 VDC/100 mA maximum
Connector
a. Input is +3.3 to 5 VDC or contact closure, active low, 30 ms minimum, 5 VDC maximum.
300 millisecond output is open collector, active low, 15 VDC/100 mA maximum.
b.
c.
Analog input is 0 to 1.0 V nominal, –1.2 to +1.2 V usable, –5.0 to +5.0 V maximum.
Table 1-5 Pump Connector Pins
Pin
Type
1
Input
2
Output
3
Output
4
Common
5
Output
6
Not used
K
Shield
Table 1-6 Detector Connector Pins
Pin
Type
1
Common
2
Input
3
Not used
4
Output
5
Input
6
Input
7
Common
8
Input
K
Shield
®
®
Foxy
R1 and Foxy
R2 Fraction Collectors
Section 1 Introduction
Description
Count: pump volume signal
Pump pause: stops pump between tubes
b
and at end of run
Valve output: control signal for optional
high flow diverter valve
Circuit common
+24 VDC Output: power for optional high
flow diverter valve
Earth ground
Description
Analog common for use with pin 8
a
Ext peak
Event: pulse delivered at tube change
Inject: starts an inject cycle1
a
Tube Advance1
Circuit common
Analog detector signal: used for peak
c
detection
Earth ground
a
b
a
1-9