Teledyne Foxy R1 Instruction Manual page 19

Hide thumbs Also See for Foxy R1:
Table of Contents

Advertisement

Foxy
Table 1-2 Specifications
Standard Diverter
Drop former (NC): 110 µL
Valve Internal Volume:
Waste (NO): 130 µL
Diverter Valve
25 psi (1.723 bar) maximum, standard and high
Inlet Pressure:
flow valves
Peak Detection:
Threshold level, peak width (slope detection), time
windows
System Control:
Operator:
Front panel control via touch screen LCD
Integrated systems:
Direct communication via Ethernet (TCP/IP) and
RS-232 serial communications.
Stored Methods:
Eight
Operating Voltage:
100–240 VAC, 50–60 Hz, 1 ampere maximum.
Mains voltage must not exceed 10% of the nomi-
nal supply voltage.
Operating
0 to 40 °C
Temperature:
32 to 104 °F
Humidity:
95% relative humidity maximum. Internal heaters
protect internal components.
Inputs:
Pump counter
advance
Outputs:
Pump pause
Wetted Materials:
Valve: PEEK and Perfluoroelastomer (FFKM)
Supplied ferrules: ETFE
Supplied valve tubing: PTFE
Supplied drain tubing: Vinyl
Typical Applications:
Prep-HPLC, peptide/protein purification, soil
extraction, density gradient fractionation.
Electrical Safety per EN 61010-1, Second Edition:
Maximum Altitude:
2000 meters
Pollution Degree:
2
Installation Category
II
a. Teledyne Isco is continually improving its products therefore specifications
are subject to change.
b. No rack is required for R1 which holds 9 bottles on the drip tray. R2 re-
quires a set of two racks which hold a total of 18 bottles.
c.
+3.3 to 5 VDC or contact closure, active low, 30 ms minimum, 5 VDC max.
d. 0 to 1.0 V nominal, –1.2 to +1.2 V usable, –5 to +5 V max.
e. Output is open collector, active low, 15 VDC/100 mA maximum
f.
300 millisecond output is open collector, active low, 15 VDC/100 mA max-
imum.
®
®
R1 and Foxy
R2 Fraction Collectors
Section 1 Introduction
a
(Continued)
c
c
, external peak
, inject
c
d
, analog peak voltage
e
, tube change event
c
, tube
.
f
.
1-7