4.5.3 Programming
Commands
Foxy
VALVE=n
This command turns the valve on or off. Setting
VALVE=0 will cause the flow to the drop former to be diverted to
waste with a three-way valve. Setting VALVE=1 will allow flow to
the drop former.
Example:
Command:
Response:
The VALVE command without an operand returns the current
valve state.
Example:
Command:
Response:
Unsupported Immediate Control Commands
The following serial commands are not supported by the fraction
collector.
• GRADIENT=n
• OSTATE=n
• OUTPUT=n
• PSPEED=n
These commands were available for earlier fraction collector
models. Do not use them with the fraction collectors.
DEFAULT=1
This command clears the current program to
defaults.
Note
To minimize accidental clearing, the characters "=1" are
required.
Example:
Command:
Response:
DELAY=n
This command sets the flow delay of the current
program. If FTYPE=1 (fraction by time), then n=0 to 359999
seconds. If FTYPE=2 or 3 (fraction by drops or external counts),
t h e n n = 0 t o 9 9 9 9 9 9 c o u n t s. Fo r e x a m p l e , t h e s t r i n g
FTYPE=1;DELAY=102 sets a flow delay of 1 minute 42 seconds.
Example:
Command:
Response:
The DELAY command without an operand returns the current
setting.
®
®
R1 and Foxy
R2 Fraction Collectors
Section 4 Serial Command Control
VALVE=1
(none)
VALVE
VALVE=1
DEFAULT=1
(none)
DELAY=102
(none)
4-9