R1 and Foxy
R2 Fraction Collectors
Section 4 Serial Command Control
4-8
RUN
The fraction collector starts, continues, or restarts a run.
Example:
Command:
Response:
R1DETECT or R2DETECT
query the two rack RFID readers that are connected to the
circuit board. R1DETECT responses with the rack identity of
Rack1 (front rack). R2DETECT responses with the rack identity
of Rack2 (back rack).
If a rack is not detected, the command will respond with
RnDETECT=0.
Example:
Command:
Response:
If a rack is detected, then the command will typically respond
within 500 milliseconds. If a rack is not detected, then the
command will timeout and typically respond within 700 milli-
seconds.
STOP
This command pauses or stops a run. Enter the command once to
pause a run in progress. Enter the command twice to completely
stop the run. This command is ignored if the unit is not in a col-
lection run.
Example:
Command:
Response:
TUBE=n
This command sets the current tube number to 'n' then moves
the drop former to it if the drop former is not at home. The 'n'
operand can be from 1 up to the maximum for the currently pro-
grammed rack type.
Example:
Command:
Response:
TUBE without an operand requests the tube number the drop
former is currently over.
Example:
Command:
Response:
RUN
(none)
(Foxy R2 only) These commands
R1DETECT
006500100075
STOP
(none)
TUBE=68
(none)
TUBE
TUBE=45