Teledyne Foxy R1 Instruction Manual page 67

Hide thumbs Also See for Foxy R1:
Table of Contents

Advertisement

Foxy
PUMP=n
This command sets the pump controlled by the
fraction collector on (PUMP=1) or off (PUMP=0).
Example:
Command:
Response:
A PUMP command without an operand returns the current
setting.
Example:
Command:
Response:
RACK=n
This command instructs the unit to move the drop
former over the home position (RACK=0) or the racks (RACK=1,
RACK=2).
Example:
Command:
Response:
A RACK command without an operand returns the current
position.
Example:
Command:
Response:
RTUBE=rnnn
This command sets the current rack number
to r and tube number to nnn and moves the drop former to it if the
drop former is not at home. nnn can be from 1 up to the maximum
for currently programmed rack type.
Note
nnn must be three characters in length.
Example:
Command:
Response:
An RTUBE command without an operand requests the rack and
tube number the drop former is currently over.
Example:
Command:
Response:
®
®
R1 and Foxy
R2 Fraction Collectors
Section 4 Serial Command Control
PUMP=1
(none)
PUMP
PUMP=1
RACK=1
(none)
RACK
RACK=1
RTUBE=1002
(none)
RTUBE
RTUBE=2035
4-7