Figure 2-11 Connections to high flow diverter valve
2.2.3 Drip Tray Drain
Connection
Foxy
Plastic valve body damage possible. While swaging the tubing
connections, always use two wrenches to prevent turning the
metal fittings in the plastic valve body. Use one wrench to hold
the inside nut to keep it from turning. Use the second wrench
to turn the outer nut.
6. Connect the other end of each tubing piece to their respec-
tive source or waste container.
7. Ensure that the diverter valve is in the highest position.
8. Move the arm and diverter valve through the full range of
motion. Ensure the In and Waste tubing do not obstruct
the movement. If necessary, secure the tubing using the
adhesive-backed clips from the accessory package.
In
5
Connect the
/
inch I.D. clear vinyl tubing to the drip tray drain
16
and route it to a suitable waste container.
Connect tubing to the drip tray drain when collecting large vol-
umes, when the installed location is at risk from any overflow,
or when collecting fluids that are flammable or can produce
flammable vapors.
To prevent fluid from pooling, drain tubing must slope down-
ward from the drip tray without loops or kinks. If necessary, ele-
vate the fraction collector by placing it on a sturdy shelf or
platform.
®
®
R1 and Foxy
R2 Fraction Collectors