Bulb Trigger {Expose During Trigger Pulse} Timing; Figure 6-5: Timing Diagram: Winx/32 Continuous Cleans; Figure 6-6: Timing Diagram: Lightfield Clean Until Trigger (Cut) - Teledyne ProEM Series System Manual

Table of Contents

Advertisement

Chapter 6

Figure 6-5: Timing Diagram: WinX/32 Continuous Cleans

Shutter (Normal)
Shutter (Preopen)
(positive polarity shown)

Figure 6-6: Timing Diagram: LightField Clean Until Trigger (CUT)

Shutter Timing Mode
(Normal)
Shutter Timing Mode
(Open Before Trigger)
6.1.5

Bulb Trigger {Expose During Trigger Pulse} Timing

When Bulb Trigger {Expose During Trigger Pulse} timing is selected, the camera
exposure is set by the External Sync input at the EXT SYNC connector. This allows an
external timing generator to control the exposure time of the camera.
In Full Frame, Frame Transfer, Kinetics, and Spectra-Kinetics modes, the transition from
the inactive state to the active state of the External Sync at the EXT SYNC connector
starts the exposure; and the transition from the active state to the inactive state ends
the Expose.
When the camera is ready to accept the external sync through the EXT SYNC
connector, the Wait for Trigger {Waiting For Trigger} (WFT) signal at the LOGIC OUT
connector is high (if WFT is the selected output signal): WFT goes HIGH immediately
after readout (or after preopen if it is active). It goes LOW when an active edge (+ or -
edge depending on your setting) occurs and, if Continuous Cleans {Clean Until Trigger}
is not enabled, the exposure begins. If continuous cleans is enabled, the camera will
check for an active edge at the EXT SYNC connector before entering a continuous
clean cycle. If none has occurred, a cycle will begin and complete. The EXT SYNC
connector is checked again to see if an active edge has occurred and exposure will
begin if it has.
Advanced Topics
Open
Open
Read Out
External Sync
Clean
CUT
Cycle
Not Reading Out
External Sync
(negative polarity shown)
t w1
and exposure
Close
Open
Close
Open
Read
Open
Close
Open
Open
Close
Open
CUT
t exp
t c
t R
First wait
Data
Second wait
stored
and exposure
Close
Open
Close
Close
Open
Close
Read
Read
Close
Open
Close
Close
Open
Close
CUT
Data
Third wait
Data
stored
and exposure
stored