Acquisition Started But Viewer Contents Do Not Update; Baseline Signal Suddenly Changes; Figure 8-2: Acquisition Display - Teledyne ProEM Series System Manual

Table of Contents

Advertisement

Chapter 8
8.2

Acquisition Started but Viewer Contents Do Not Update

In LightField, live data is normally displayed in the Experiment workspace viewer as it is
being acquired (Preview or Acquire mode). If the viewer is not being updated and
acquisition is occurring, check to see if there is a filename in the top row of the viewer:
in
Figure
filename, click on the camera icon. The file data will be cleared from the viewer and the
live data will then be displayed.

Figure 8-2: Acquisition Display

8.3

Baseline Signal Suddenly Changes

A change in the baseline signal is normal if the temperature, gain, or speed setting has
been changed. If this occurs when none of these settings have been changed, there
may be excessive humidity in the camera vacuum enclosure. Turn off the camera and
contact Teledyne Princeton Instruments Customer Support. Refer to
Information
Troubleshooting
8-2, the filename 2010 May 11 13_13_49.raw is displayed. If there is a
on page 160 for complete information.
125
Contact

Advertisement

Table of Contents
loading

This manual is also suitable for:

Proem:512bProem:512bkProem:1024bProem:1600

Table of Contents