Chapter 6
6.1.2
Free Run {No Response}
In this mode, there is no external triggering {or the camera is ignoring external triggers}
and all settings are read from the setup parameters, making the duration of each
exposure time constant and the interval times between exposures constant. See
Figure
Figure 6-2: Free Run Timing Diagram
Exposure
Read Out
6.1.3
External Sync {Readout Per Trigger}
In this mode, each frame in a sequence requires a trigger. Each frame is exposed for the
length of time entered into the software and is then read out. If a trigger arrives during
the exposure-readout of the previous frame, it is ignored. See
Figure 6-3: External Sync Timing Diagram
Trigger In
Exposure
Read Out
For a sequence of one frame, bulb mode and trigger-first mode are the same. The
shaded areas
Advanced Topics
6-2.
#1
#1
#1
#1
denote the idle time between exposures.
#2
#3
#2
Ignored
#2
#3
#2
#4
#3
#4
Figure
6-3.
#4
#3
#4
91