Data Overrun Due To Hardware Conflict Message; Figure 8-5: Data Overrun Due To Hardware Conflict Dialog - Teledyne ProEM Series System Manual

Table of Contents

Advertisement

Chapter 8
8.8

Data Overrun Due to Hardware Conflict message

Figure 8-5: Data Overrun Due to Hardware Conflict dialog

If this dialog appears when you try to acquire a test image, acquire data, or run in focus
mode, check the CCD array size and then check the DMA buffer size. A large array (for
example, a 2048 x 2048 array), requires a larger DMA buffer larger setting than that for a
smaller array (for example, a 512 x 512 array).
To change the DMA buffer setting:
1.
Note the array size (on the Setup
Acquisition
2. Open Setup
3. Increase the DMA buffer size to a minimum of 32 MB (64 MB if it is currently 32 MB
or 128 MB if it is currently 64 MB), click on OK, and close the WinX/32 application
(WinView/32 or WinSpec/32).
4. Reboot your computer.
5. Restart the WinX/32 application and begin acquiring data or focusing. If you see the
message again, increase the DMA buffer size.
Troubleshooting
Experiment Setup
Environment
Environment dialog.
Hardware
Controller/CCD tab or the
Main tab Full Chip dimensions).