Configuring Software Parameters In Winx/32 - Teledyne ProEM Series System Manual

Table of Contents

Advertisement

104
This procedure assumes:
6.4.5.1

Configuring Software Parameters in WinX/32

Configure the software parameters as follows:
ProEM System Manual
You have already set up your system in accordance with the instructions in the
previous chapters.
You have read the previous sections of this chapter.
You are familiar with the application software.
The system is being operated in imaging mode.
The target is a sharp image, text, or a drawing that can be used to verify that the
camera is seeing and can be used to maximize focus.
You are only illuminating a portion of the CCD as illustrated in
You are using either WinX/32 or LightField. WinX/32 parameter configuration is
presented first; LightField, second.
NOTE:
The following procedure is based on WinView/32. You will
need to modify it if you are using a different application.
Basic familiarity with the WinView/32 software is assumed. If
this is not the case, you may want to review the software
manual or have it available while performing this procedure.
Controller/Camera Tab (Setup
These parameters should be set automatically to the proper values for your
system.
— Controller type: This information is read from the camera.
— Camera type: This information is read from the camera.
— Shutter type: Custom (System dependent).
— Readout Mode: Available modes are read from the camera. Select Kinetics.
— Window Size: Enter the number of exposed rows.
Cleans Tab (Setup
Hardware): Automatically set.
Detector Temperature (Setup
The default temperature setting is read from the camera. When the array
temperature reaches the set temperature, the Detector Temperature dialog will
report that the temperature is LOCKED.
Experiment Setup
Main Tab (Acquisition
— Exposure Time: 100
— Accumulations & Number of Images: 1
Experiment Setup
ROI Tab (Acquisition
Use this function to define the region of interest (ROI).
— Imaging Mode: Select this mode if you are running WinSpec.
— Clicking on Full loads the full size of the chip into the edit boxes.
Experiment Setup
Timing Tab (Acquisition
— Timing Mode: Free Run, Single Trigger, or Multiple Trigger
— Shutter Control: Normal
— Safe Mode vs. Fast Mode
Safe Mode must be selected for Kinetics mode operation.
Hardware)
Detector Temperature...)
Experiment Setup...):
s
Experiment Setup...)
Figure
6-12.
Experiment Setup...):