Chapter 5
5.4.6.2

Full Frame (Sequential) Mode for Frame-Transfer EMCCD

Full Frame mode allows you to expose the array for the exposure time specified in the
software and is similar in performance to a normal, full-frame CCD device. The
operational sequence for this mode is:
1.
Clearing the CCD.
2. Exposing for the specified exposure time.
3. Shifting the image from the sensor area to the frame-transfer area.
4. Reading out the CCD.
See
Figure

Figure 5-22: Functional Array Structure Diagram: Frame-Transfer EMCCD

For standard, high dynamic range
Step 1 through Step 4 are repeated for each frame in a sequence. Step 1 and Step 3,
clearing the CCD and shifting the image, are usually very short and do not impact the
frame rate.
Operation
5-22.
Extended Multiplication Register
Readout
Amplifier
Multiplication Gain
For high speed, low light level
applications
Normal
applications
Readout
Amplifier
Output or Sensor Node
In WinX/32, Full Frame mode can be selected from Hardware Setup
Controller/Camera tab.
In LightField, it is selected on the Sensor expander.
NOTE:
Since the software you are using may show the settings
differently, refer to the software documentation for
accurate information.
Output or Sensor Node
Standard Serial Register
Frame-transfer Area
Sensor Area
High Voltage Clock
Normal Voltage Clock