SECTION 12—CENTER MOUNT
FOOTRESTS
After any adjustments, repair or service and before use, make sure all attaching
hardware is tightened securely - otherwise injury or damage may result.
While the wheelchair is moving, minimum ground clearance for the front rigging is
three inches. If the wheelchair is not moving, the front rigging MUST maintain a
minimum of one inch ground clearance - otherwise personal injury and damage may
result.
Before performing any maintenance, adjustments or service, always turn the wheel-
chair power off, otherwise injury or damage may result.
NOTE: For complete operating information on Invacare footrests, refer to the appropriate Owner's
Manual listed in Reference Documents on page 8.

Removing/Installing the Center Mount Footrest

NOTE: For this procedure, refer to FIGURE 12.1 on page 54.
Removing
1. Remove the rigging pivot pin that secures the center mount footrest to the seat frame
mounting bracket.
2. Hold the center mount footrest with one hand and engage the release lever with the
other while simultaneously pulling the center mount footrest out of the ratchet
housing.
Installing
1. Engage the release lever with one hand, hold the center mount footrest with the other,
and insert the center mount footrest into the ratchet housing.
2. Reinstall the rigging pivot pin to secure the center mount footrest to the seat frame
mounting bracket.
Part No 1125031
SECTION 12—CENTER MOUNT FOOTRESTS
WARNING
53
Formula
PTO Plus Powered Seating