Invacare Formula PTO Plus Service Manual page 33

Powered seating on tdx wheelchairs pronto m71 wheelchairs pronto m91 wheelchairs
Hide thumbs Also See for Formula PTO Plus:
Table of Contents

Advertisement

7
E07
Remote Fault
8
E08
Controller Fault
9
E09
Communications
Fault
10
E10
General Fault
11
E11
Incompatible/
incorrect Remote
Part No 1125031
SECTION 10—SAFETY INSPECTION/TROUBLESHOOTING
00
Local SR Fault (CPU,
EEPROM, etc.)
01
Joystick fault at the remote
02
Speed pot fault at the remote
00
Controller fault
01
RAM fault
02
ROM fault
03
CPU fault
04
EEPROM fault
05
Watchdog fault
06
Stack fault
07
Software fault
08
Power-up testing fault
09
Relay fault or precharge fault
10
Bridge fault or disable all fault
11
Electronics fault: Thermistor
12
Calibration setting fault
00
Remote connection lost
01
Low communication mode
00
General fault
00
Incompatible/incorrect
Remote
33
Check the communications bus,
connections and wiring. Replace the
remote.
Check connections and wiring. Replace
power module.
Check connections and wiring. Replace
Bus cable.
Check all connections and wiring. Contact
Invacare Technical Service.
Wrong type of remote connected. Ensure
the branding of the joystick matches that
of controller unit.