Risk of Death or Serious Injury
NEVER operate the wheelchair or elevate/lower the seat while in any tilted/
reclined/back angle position over 20° relative to the vertical position. If the drive
lock-out does not stop the wheelchair from operating or the seat from elevating/
lowering in a tilt/recline/back angle position over 20° relative to vertical, DO NOT
operate the wheelchair or elevate/lower the seat. The drive lock-out settings MUST
be adjusted. Refer to Adjusting/Replacing the Drive Lockout Sensor on page 134.
The wheelchair user MUST have a clear line of sight to drive safely. On initial chair
delivery and after adjusting the back angle, drive lock-out switch tilt system or
recline system, tilt and recline the seat back to the farthest driving position
IMMEDIATELY before drive lock-out engages and ensure there is a clear line of
sight present in which to drive the wheelchair. If a clear line of sight is not present,
have the back angle repositioned or readjust the lockout angle such that safe driving
with a clear line of sight is achieved.
5. Repeat STEP 1 to determine if the adjustment achieved the desired drive lockout
angle.
6. Perform one of the following:
• Drive Lockout Angle is Correct - Proceed to STEP 7.
• Drive Lockout Angle is Incorrect - Repeat STEPS 1-5 until the drive lockout angle
is correct.
7. Test the tilt function and ensure the drive lockout works at no more than 20° of tilt.

Adjusting/Replacing the Tilt Sensor

The cam switch harness contains the tilt sensor. There is no adjustment to this switch.
Refer to Removing/Installing the Cam Harness Cable on page 114 for replacement
instructions.

Removing/Installing the Sanode

NOTE: For this procedure, refer to FIGURE 17.3 on page 122.
NOTE: Use the sanode to operate the single actuator seating system using the Mk6i MPJ+ joystick.
Removing
1. Remove the tie-wraps securing the sanode and cable to the wheelchair.
2. Disconnect the network connector. Refer to Connecting/Disconnecting Network
Connectors on page 118.
Part No 1125031