Verifying the System
Method 1: Using a Signal Generator
Preparing the Software
Do This
1 Start the application program by going to the
2 Create a new project called
3 Click Configuration and then add the receiver
4 Click Collection and select the Spectrum VFP
5 Click Maximize button in the upper-right corner
6 Connect the signal generator output to the RF
7 Turn on the signal generator.
8 Set the measurement software as follows:
Agilent Wireless Solutions
Start menu, selecting Programs, Agilent E74xx.
Verification Test 1 (External generator).
hardware.
(Virtual Front Panel)
of the main window.
input of the receiver.
Band: Uplink
Frequency: See
Table 4-1 on page 140
Span: 1 MHz
Averaging: Running
Averages: 32
Additional Information
The software starts.
We recommend you use the automatic
configuration wizard for this process. Refer to
"Automatic Configuration" on page
Refer to
"Hardware Installation Tips" on page 21
The display size increases to fill the available
space.
You should see a signal trace showing a single
spectral peak in the middle of the display.
Figure 4-2 page 140
shows a example trace.
23.
139