Olympus BF-UC180F Instructions Manual page 42

Bf type evis exera ii ultrasonic bronchofibervideoscope
Hide thumbs Also See for BF-UC180F:
Table of Contents

Advertisement

3.
Inspect and verify that the base of the valve is in contact with the suction
cylinder. If a gap still exists between the valve and the suction cylinder, the
valve is not properly attached and patient debris or fluid may leak or spray
from the gap (see Figure 3.11).
Proper attachment
Figure 3.11
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
No gap
Sometimes the suction valve will click before it is fully seated
in the suction cylinder. Press the suction valve down firmly to
ensure that it is fully seated in the suction cylinder and
visually inspect the connection to ensure that no gap exists.
Chapter 3 Preparation and Inspection
Improper attachment
Gap
35