Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
Flushing detergent solution into the
instrument/suction/balloon channels
110
1.
Disconnect the suction tube from the suction cleaning adapter. Completely
immerse the endoscope and suction cleaning adapter in detergent solution.
2.
Attach the 30 cm
3.
Withdraw the plunger of the syringe to fill the suction cleaning adapter and
the channel with the detergent solution (see Figure 7.21).
4.
Attach the syringe to the irrigation port and withdraw the plunger of the
syringe to fill the balloon channel with the detergent solution (see Figure
7.22).
5.
With the endoscope and suction cleaning adapter completely immersed in
detergent solution, disconnect the suction cleaning adapter and syringe.
Do not pull the tubes of the suction cleaning adapter when
disconnecting the suction cleaning adapter from the
endoscope.
Suction opening
Suction cylinder cap
Figure 7.21
Irrigation port
Figure 7.22
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
3
(30 ml) syringe to the suction opening onto the adapter.
Irrigation port
Port cap