Chapter 3 Preparation and Inspection
Inspection of the single use adapter biopsy valve (MAJ-1414)
30
The single use adapter biopsy valve (MAJ-1414) is provided
in a sterile condition. Do not open the package until you are
ready to use it.
Do not use the instrument after the expiration date displayed
on the sterile package. Doing so may pose an
infection-control risk.
Do not attempt to reuse or sterilize the single use adapter
biopsy valve (MAJ-1414). Doing so could pose an
infection-control risk or cause equipment damage.
This single use adapter biopsy valve (MAJ-1414) is designed
for use with the single use aspiration needle
(NA-201SX-4022).
1.
Inspect the sterile package for tears, inadequate sealing, or water damage.
If the sterile package shows any irregularity, the sterile condition of the
instrument may have been compromised. Use a spare instead.
2.
Inspect the biopsy valve for cracks or other damage (see Figure 3.6).
Housing
Figure 3.6
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Slit