Olympus BF-UC180F Instructions Manual page 136

Bf type evis exera ii ultrasonic bronchofibervideoscope
Hide thumbs Also See for BF-UC180F:
Table of Contents

Advertisement

 Steam sterilization (autoclaving)
After cleaning as described in Section 7.5, "Manual cleaning" on page 102,
steam sterilize (autoclave) following the procedure given below.
1.
Seal the individual parts or equipment separately in packages appropriate
for steam sterilization (autoclaving) according to your hospital's protocol.
2.
Steam sterilize the packages according to the recommended steam
sterilization (autoclaving) exposure parameters as described in Section 6.7,
"Steam sterilization (autoclaving) of accessories" on page 81 and the
sterilizer manufacturer's instructions.
3.
Following steam sterilization (autoclaving), let all components gradually cool
down to room temperature.
4.
Store the components following the instructions given in Chapter 9, "Storage
and Disposal".
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
Before taking the equipment package out of the autoclave, let
it cool down to room temperature. Otherwise, it may cause
burns.
Effective sterilization will not be possible if items are packed
tightly together in the autoclave; always pack items loosely.
Inspect each equipment package for openings, tears or other
damage. If the equipment package is open or damaged, seal
the equipment in new package and resterilize it as described
below.
Allow the packages to dry within the autoclave, using the
autoclave's drying cycle (if available) or by opening the door
of the autoclave and allowing the packages to air-dry.
Handling a wet package can compromise its sterility.
Sudden changes in temperature may damage the
instruments.
129