Olympus BF-UC180F Instructions Manual page 132

Bf type evis exera ii ultrasonic bronchofibervideoscope
Hide thumbs Also See for BF-UC180F:
Table of Contents

Advertisement

21.
Using a sterile, lint-free cloth, thoroughly wipe and dry all external surfaces.
High-level disinfection
1.
Fill a basin with disinfectant solution at the temperature and concentration
recommended by the disinfectant manufacturer. Use a basin that is deep
enough to completely immerse all equipment.
2.
Immerse all equipment in the disinfectant solution.
3.
Using a lint-free cloth and/or 30 cm
surfaces with the disinfectant solution to remove all air bubbles.
4.
Using the 30 cm
them while they are immersed. Ensure that all air bubbles are expelled.
5.
Using the 30 cm
valve with disinfectant solution while the valve is immersed. Ensure that all
air bubbles are expelled (only for reusable biopsy valve (MD-495)).
6.
Rub the bristles of the channel cleaning brushes and channel-opening
cleaning brush to remove all air bubbles.
7.
Using the 30 cm
adapter thoroughly with disinfectant solution to remove all air bubbles while
it is immersed in the disinfectant solution.
8.
Cover the basin with a tight fitting lid to minimize the release of disinfectant
vapor. Soak all equipment at the temperature and concentration
recommended by the disinfectant manufacturer.
9.
Remove the suction cleaning adapter from the disinfectant solution. Hold
the end of the adapter and tilt it for expelling the disinfectant solution
remaining in its adapter.
10.
Remove all equipment from the disinfectant solution.
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
Completely remove all air bubbles from all equipment,
including inner channels. Air bubbles remaining on the
equipment prevent proper disinfection.
All disinfection steps should be performed with all equipment
completely immersed. If any part of the equipment is not
immersed completely, disinfectant solution may not
adequately contact all surfaces.
3
(30 ml) syringe, flush the interior and recessed parts of
3
(30 ml) syringe, flush the recessed parts of the biopsy
3
(30 ml) syringe, flush the inside of the suction cleaning
3
(30 ml) syringe, wipe and/or flush all
125