Olympus BF-UC180F Instructions Manual page 98

Bf type evis exera ii ultrasonic bronchofibervideoscope
Hide thumbs Also See for BF-UC180F:
Table of Contents

Advertisement

 Inspection of the single use combination cleaning brush
(BW-411B)
1.
Remove the brush from its packaging just prior to use (see Figure 7.12).
2.
Confirm that the channel cleaning brush part and the tip at the distal end
are securely attached.
3.
Check the channel cleaning brush and the channe-opening cleaning brush
parts for loose or missing bristles.
4.
Check the bristles of the channel cleaning brush and the channel-opening
cleaning brush parts for any damage. If the bristles are crushed, gently
straighten them with your fingertips.
5.
Check the shaft for bends, scratches, and other damage. Do not reprocess
the single use combination cleaning brush prior to use. The brush may be
damaged.
Handle
Figure 7.12
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
Brush head
Tip
Shaft
Packaging
91