Chapter 2 Instrument Nomenclature and Specifications
22
Connecting hook
Chain
Figure 2.1
To ensure that you do not forget to attach the water-resistant
cap, it is recommended that you connect it to the connecting
ring of endoscope connector using the cap chain.
1.
Confirm that the cap chain is free from cracks, flaws, wear, deformation, or
other damages (see Figure 2.1).
2.
Align the notch on the connecting plate with the pin on the venting connector
of the water-resistant cap (MH-553, see Figure 2.2).
3.
Place the connecting plate over the venting connector (see Figure 2.2).
4.
Confirm that the connecting plate is securely attached to the foot of the
venting connector and can be smoothly rotated.
5.
Attach the connecting hook to the connecting ring (see Figure 2.3).
Connecting plate
Venting connector
Water-resistant cap
Figure 2.2
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Connecting plate
Notch
Pin
Notch