Inspection of the single-use biopsy valve (MAJ-210)
1.
Inspect the sterile package for tears, inadequate sealing, or water damage.
If the sterile package shows any irregularity, the sterile condition of the
instrument may have been compromised. Use a spare instead.
2.
Inspect the biopsy valve for cracks or other damage (see Figure 3.7).
Figure 3.7
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
•
The single-use biopsy valve (MAJ-210) is provided in a
sterile condition. Do not open the package until you are ready
to use it.
•
Do not use the instrument after the expiration date displayed
on the sterile package. Doing so may pose an
infection-control risk.
•
Do not attempt to reuse or sterilize the single-use biopsy
valve (MAJ-210). This could pose an infection-control risk or
cause equipment damage.
•
Do not use the single-use biopsy valve (MAJ-210) with the
single use aspiration needle (NA-201SX-4022). Please use
the single use adapter biopsy valve (MAJ-1414). And refer to
the instruction manual for the single use aspiration needle
(NA-201SX-4022).
The single-use biopsy valve (MAJ-210) is designed for use
with BF-30/40/60/200/240/260/160/and 180 series
endoscopes.
Housing
Chapter 3 Preparation and Inspection
Slit