Olympus BF-UC180F Instructions Manual page 91

Bf type evis exera ii ultrasonic bronchofibervideoscope
Hide thumbs Also See for BF-UC180F:
Table of Contents

Advertisement

Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
Reprocessing equipment parts and functions
84
For inspection of other equipment than that mentioned below, refer to the
instruction manual for the equipment being used.
 Water-resistant cap (MH-553)
The water-resistant caps are attached to the videoscope scope cable connector
and the ultrasonic cable connector on the endoscope to protect the connectors
from water penetration during reprocessing. For leakage testing, the venting
connector on the water-resistant cap must be connected to the leakage tester
(MB-155) (see Figure 7.2).
Groove
Seal
Figure 7.2
 Suction cleaning adapter (MAJ-222)
The suction cleaning adapter is used to aspirate reprocessing fluids from the
distal end of the endoscope through the instrument channel and the suction
channels (see Figure 7.3).
Suction opening
Port cap
Figure 7.3
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Venting connector
Suction
cylinder cap