ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Chapter 8 Cleaning and Disinfection Equipment
When cleaning and disinfecting an endoscope in an
endoscope reprocessor, use connectors that are specifically
compatible with the endoscope model. Otherwise, insufficient
cleaning and disinfection or sterilization of the endoscope
may occur, which could pose an infection-control risk to the
patient and/or operators performing the next procedure with
the endoscope. The applicable connectors for each
endoscope model should be listed in the instruction manual
for the endoscope reprocessor. If your endoscope model is
not listed in that instruction manual, please contact the
manufacturer of the endoscope reprocessor.
If using the endoscope reprocessor recommended by
Olympus, be sure to attach the appropriate
cleaning/connecting tube and retaining rack (see Table 8.1).
Otherwise, insufficient cleaning and disinfection of the
endoscope may pose an infection-control risk to the patient
and/or operators performing the next procedure with the
endoscope.
Cleaning/connecting tube
MAJ-50, MAJ-1190
MAJ-831, MAJ-1191
MAJ-1513, MAJ-1515
Retaining rack
MAJ-840
MAJ-840