Nonsterile water rinse and alcohol flush
1.
Fill a small basin with clean water.
2.
Immerse all equipment in clean water and follow all procedures described in
"Sterile water rinse" on page 126.
3.
Fill a small basin with 70% ethyl or isopropyl alcohol.
4.
Immerse all equipment in the alcohol.
5.
Gently agitate the equipment in the alcohol.
6.
Using the 30 cm
equipment with alcohol to remove all air bubbles.
7.
Rub the bristles of the channel cleaning brushes and channel-opening
cleaning brush to remove all air bubbles.
8.
Attach the 30 cm
the adapter with 30 cm
9.
Remove the suction cleaning adapter from the alcohol and remove all
alcohol from it.
10.
Remove all equipment from the alcohol.
11.
Using a sterile, lint-free cloth, thoroughly wipe and dry all external surfaces.
12.
Dry all equipment.
13.
Store the components following the instructions given in Chapter 9, "Storage
and Disposal".
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
3
(30 ml) syringe, flush the interior and recessed parts of all
3
(30 ml) syringe to the suction cleaning adapter and flush
3
(30 ml) of alcohol to remove all air bubbles.
127