SECTION 6—PANEL ASSEMBLY

Replacing Accumulator Assembly

DANGER
To prevent electrical shock, ALWAYS disconnect compressor from electrical
outlet before servicing.
NOTE: For this procedure, refer to FIGURE 6.5 on page 33.
Tools required for this procedure are:
• Phillips screwdriver
Removing Accumulator Assembly
1. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 25.
2. Remove the control panel. Refer to Removing/Installing Control Panel on page 29.
3. Remove the two mounting screws securing the accumulator assembly to the main
brackets.
4. Cut the two tie-wraps securing the two fittings to the oxygen hoses.
5. Remove the existing accumulator assembly.
Installing Accumulator Assembly
1. Install the accumulator assembly to the main brackets with two mounting screws.
2. Connect the two fittings to the oxygen hoses.
3. Secure the two fittings to the oxygen hoses with two tie-wraps.
4. Install the control panel. Refer to Removing/Installing Control Panel on page 29.
5. Install the cabinet. Refer to Removing/Installing the Cabinet on page 25.
Oxygen Hose
Fitting
Main Bracket
Fitting
Tie Wrap
Main Bracket
Mounting Screw
Accumulator
Assembly
FIGURE 6.5 Replacing Accumulator Assembly
33
®
Part No 1154316
HomeFill
II AW Compressor