SECTION 12—WIRING ASSEMBLIES
7. Disconnect the following connectors:
A. Two connectors to the high pressure switch (Detail "B").
B. Two connectors to the circuit board (Detail "C").
C. Two connectors to the On/Off switch (Detail "C").
D. Connector to the motor (Detail "D").
8. Note the orientation of the two connectors to the exhaust fan (Detail "D").
9. Disconnect the connectors from the exhaust fan.
10. Reverse STEPS 1-9 to install the new HFII assembly wire harness.
DETAIL "A"
HF II Assembly
Wire Harness
Power Inlet
Connectors
DETAIL "C"
On/Off
Switch
Connectors
FIGURE 12.4 Replacing HFII Assembly Wire Harness
®
HomeFill
II AW Compressor
Hour Meter
Connectors
Motor
Connector
Circuit
Board
Connectors
84
DETAIL "B"
DETAIL "D"
High Pressure
Switch
Connectors
Exhaust Fan
Connectors
Part No 1154316