Removing/Installing Control Panel

NOTE: For this procedure, refer to FIGURE 6.2.
Tools required for this procedure are:
• Phillips screwdriver
1. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 25.
2. Remove the two mounting screws securing the top of the control panel (Detail "A").
3. Loosen the two mounting screws securing the bottom of the control panel (Detail "A").
4. Slowly pull the control panel back to expose the circuit board and its connections.
5. Disconnect the three wiring harnesses from the three connectors on the circuit board
(Detail "B").
DO NOT disconnect the opaque oxygen sensor tube from the sensor as this may
damage the circuit board.
6. Disconnect the opaque oxygen sensor tube from the barbed fitting opposite the sensor
(Detail "B").
7. Note the location of the two wire connectors of the power switch for later replacement.
8. Disconnect the two wire connectors from the power switch (Detail "B").
9. Disconnect bleed resistor from the control panel (Detail "C").
10. Remove the control panel.

Installing

1. Connect bleed resistor to the control panel (Detail "C").
2. Connect the two wire connectors to the power switch (Detail "B") in the orientation
noted previously.
3. Connect the opaque oxygen sensor tube to the barbed fitting on the opposite end of
the sensor (Detail "B").
4. Connect the three wiring harnesses to the three connectors on the circuit board (Detail
"B").
5. Secure the bottom of the control panel with two mounting screws. Securely tighten.
6. Secure the top of the control panel with two mounting screws. Securely tighten.
7. Install the cabinet. Refer to Removing/Installing the Cabinet on page 25.