SECTION 5—CABINET ASSEMBLY
Removing/Installing the Cabinet
To prevent electrical shock, ALWAYS disconnect HomeFill from electrical outlet
before servicing.
NOTE: For this procedure, refer to FIGURE 5.1.
Tools required for this procedure are:
• Phillips screwdriver
Removing
1. Unplug the power cord from the electrical outlet.
2. Remove cylinder (if attached). Refer to Disconnecting Cylinder From Compressor on
page 22.
3. Remove the four mounting screws that secure the cabinet assembly to the base
assembly.
4. Lift the cabinet straight up off of the base assembly.
Installing
1. Position the cabinet on the base assembly as shown.
2. Secure the cabinet to the base assembly with four mounting screws. Securely tighten.
3. Reattach cylinder. Refer to Connecting the Cylinder to the Compressor on page 20.
4. Plug power cord into electrical outlet.
Cabinet
Mounting Screws
Part No 1154316
DANGER
FIGURE 5.1 Removing/Installing the Cabinet
SECTION 5—CABINET ASSEMBLY
Cabinet
Base Assembly
Mounting Screws
25
REAR VIEW
®
HomeFill
II AW Compressor