Reassembly - Invacare HomeFillII AW Service Manual

Table of Contents

Advertisement

Pump Assembly
Motor Pulley
DETAIL "B"
Motor
Bushing
Mounting
Screw

Reassembly

NOTE: For this procedure, refer to FIGURE 9.1 through FIGURE 9.6.
1. Position the washers on the pump assembly. Refer to Detail "C" in FIGURE 9.6.
2. Position the motor on the pump assembly. Refer to Detail "B" in FIGURE 9.6.
3. Secure the two washers and motor to the pump assembly with two mounting screws
and two bushings. Torque mounting screws to 60-70 in-lbs (6.78-7.9 Nm). Refer to
Detail "B" in FIGURE 9.6
Part No 1154316
Timing
DETAIL "A"
Belt
Pump
Assembly
Bushing
Mounting
Screw
FIGURE 9.6 Motor Replacement
57
SECTION 9—MOTOR REPLACEMENT
Large Pulley
DETAIL "C"
Washers
HomeFill
Pump
Assembly
®
II AW Compressor