Invacare HomeFillII AW Service Manual page 56

Table of Contents

Advertisement

SECTION 9—MOTOR REPLACEMENT
23. Remove the six mounting screws that secure the pump/motor assembly to the base
(FIGURE 9.5).
TOP VIEW WITH CONTROL PANEL ASSEMBLY REMOVED
Base
Mounting
Screws
24. Carefully place the pump assembly upside down on the table.
25. Remove the timing belt by gently pulling the belt away from the large pulley and
motor pulley. Refer to Detail "A" in FIGURE 9.6.
26. Rotate pump assembly 180° so that the motor is facing up.
27. Remove the two mounting screws, two bushings, and two washers that secure the
motor to the pump assembly. Refer to Detail "B" and "C" in FIGURE 9.6.
28. Remove the motor from the pump assembly.
NOTE: Retain the two mounting screws, two bushings, and two washers for reassembly.
®
HomeFill
II AW Compressor
Mounting Screws
FIGURE 9.5 Motor Replacement
56
Motor/Pump Assembly
Mounting
Screws
Part No 1154316

Hide quick links: